Z-1479-2014 Class II Terminated

Recalled by Hospira Inc. — Lake Forest, IL

FDA device recall Z-1479-2014 was initiated by Hospira Inc. on Feb 25, 2014 and is designated Class II. Reason for recall: Hospira identified an incorrect lower lid (set component) had been supplied and was being used during the manufacturing process of list number 11241-03, a HemoSet containing a burette and blood… The recall status is terminated (terminated Nov 28, 2016). Affected quantity: 3,893 sets.

Recall Details

Product Type
Devices
Report Date
Apr 30, 2014
Initiation Date
Feb 25, 2014
Termination Date
Nov 28, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3,893 sets

Product Description

LifeShield LATEX-FREE HemoSet 100 mL Burette I.V. PlumSet, Nonvented, 105 inch with Prepierced Reseals, OPTION-LOK and Inline Dual Channel Cassette. For administration of blood and blood bags for use with Plum Series Infusers.

Reason for Recall

Hospira identified an incorrect lower lid (set component) had been supplied and was being used during the manufacturing process of list number 11241-03, a HemoSet containing a burette and blood filter, intended for infusion of blood/blood products. In a gravity delivery, the correct lower lid dispenses 15 drops per mL and the incorrect lower lid found dispenses 10 drops per mL.

Distribution Pattern

US Distribution including the states of FL, IA,IN, KS, KY, ME, MO, MT, PA, TN, TX, UT, VA and WA.

Code Information

Product List Number: 112410403; Lot Numbers: 280055H, 341005H

Source