Z-1499-2016 Class II Terminated

Recalled by Leica Microsystems, Inc. — Buffalo Grove, IL

FDA device recall Z-1499-2016 was initiated by Leica Microsystems, Inc. on Jan 11, 2016 and is designated Class II. Reason for recall: An electronic component of the Leica M525 OH4 and Leica M720 OH5 surgical microscopes with 400W illumination could potentially overheat, resulting in smoke and smell emanating from the ventilation… The recall status is terminated (terminated Apr 24, 2017). Affected quantity: 381 units.

Recall Details

Product Type
Devices
Report Date
Apr 27, 2016
Initiation Date
Jan 11, 2016
Termination Date
Apr 24, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
381 units

Product Description

Leica Microsystems (Schweiz) Ag. The common name of the device is surgical microscope. It has been assigned the product code FSO by the FDA and is classified as a Class I medical device. Intended for improving the visibility of objects through magnification and illumination.

Reason for Recall

An electronic component of the Leica M525 OH4 and Leica M720 OH5 surgical microscopes with 400W illumination could potentially overheat, resulting in smoke and smell emanating from the ventilation openings of the illumination units located opposite and away from the surgeon / patient side of the surgical microscope.

Distribution Pattern

Worldwide Distribution-US (nationwide) including Puerto Rico and the states of MA, CA, SC, NJ, TX, CO, FL, PA, NY, MO, WI, NC, UT, ID, AZ, MI, KY, CT, OR, , IL, AL, MN, MT, OH, NE, GA, IN, SD, TN, OK, AK, KS, NV, NM, , ND, VA, WA, WY and MD, and the countries of China, Belgium, France, Germany, Great Britain, Italy, Norway, Portugal, Spain, Sweden, Switzerland, India, Japan, Argentina, Chile, Colombia, Mexico, Peru, Belarus, Egypt, Greece, Morocco, Poland, Qatar, Romania, Russian Fed., Saudi Arabia, South Africa, Turkey, Canada, , Australia, Malaysia and Taiwan.

Code Information

OH5-51313; OH4-112214; OH5-49112; OH5-67914; OH4-101213; OH5-20238; OH4-91913; OH5-43312; OH5-54113; OH5-64514; OH4-103014; OH4-106514; OH4-87112; OH4-95413; OH4-94213; OH4-106014; OH4-105514; OH4-94813; OH4-102514; OH4-91113; OH4-87312; OH4-106714; OH5-62113; OH5-46112; OH5-50512; OH4-85912; OH4-81512; OH5-61013; OH5-53913; OH4-95013; OH4-91413; OH4-110614; OH4-94613; OH4-104714; OH4-87212; OH4-85712; OH5-50212; OH4-96213; OH4-88913; OH4-89613; OH5-55113; OH4-99413; OH5-48112; OH4-90713; OH5-66314; OH5-68314; OH4-109414; OH4-83912; OH5-48012; OH4-101313; OH4-86812; OH5-60113; OH5-60313; OH5-37311; OH4-86612; OH5-62913; OH4-95813; OH4-90613; OH4-103614; OH5-48212; OH5-48312; OH5-58313; OH5-53513; OH5-63313; OH4-92613; OH5-43812; OH5-56713; OH5-68114; OH4-81412; OH4-98013; OH5-33911; OH5-61913; OH5-52813; OH4-110714; OH5-46312; OH5-51113; OH4-106814; OH4-112714; OH5-63514; OH5-39012; OH5-63013; OH4-111614; OH4-97013; OH4-97613; OH4-107414; OH4-109514; OH4-109614; OH4-92513; OH4-104614; OH4-93113; OH4-85812; OH4-105414; OH4-99913; OH4-91813; OH4-94913; OH4-107914; OH4-83812; OH4-100413; OH4-107214; OH5-41012; OH5-56013; OH4-85612; OH4-99713; OH4-97513; OH5-50312; OH4-109814; OH5-52013; OH5-61413; OH5-61513; OH5-62613; OH5-68614; OH5-51413; OH4-83112; OH4-108114; OH5-68214; OH4-96513; OH5-51513; OH4-100613; OH4-88213; OH5-58813; OH4-91613; OH5-53713; OH5-53813; OH4-98513; OH4-97713; OH5-36411; OH4-106414; OH5-63213; OH5-63413; OH4-100313; OH4-105914; OH5-67214; OH4-81612; OH4-108314; OH5-46612; OH5-46512; OH5-51913; OH4-96713; OH4-103314; OH4-83212; OH4-87012; OH4-95513; OH4-95713;OH5-35611; OH4-86512; OH4-104314; OH4-104414; OH4-110814; OH4-92413; OH4-100013; OH5-54913; OH5-56213; OH4-102414; OH5-39512; OH5-43712; OH4-98713; OH4-98813; OH4-98913; OH5-43912; OH5-39812; OH5-49012; OH5-39112; OH5-48712; OH5-50813; OH5-35811; OH5-55713; OH4-92713; OH5-58213; OH5-50713; OH4-102214; OH4-112614; OH4-96913; OH4-85512; OH4-92213; OH4-100513; OH4-101113; OH4-20446; OH4-20977; OH4-21557; OH4-22207; OH4-22378; OH4-24019; OH4-69311; OH4-70811; OH4-71411; OH143+684-74711; OH4-76212; OH4-76612; OH4-79312; OH4-79912; OH4-81312; OH4-82512; OH4-99213; OH5-20218; OH5-20328; OH5-20558; OH5-20788; OH5-20808; OH5-20818; OH5-20828; OH5-20858; OH5-21108; OH5-21128; OH5-21379; OH5-21679; OH5-22410; OH5-24410; OH5-20618

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1499-2016
openFDA file date
Oct 27, 2026
Loaded by RegDataLab
Oct 29, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.