Z-1508-2022 Class II Ongoing

Recalled by Alere San Diego, Inc. — San Diego, CA

Recall Details

Product Type
Devices
Report Date
August 17, 2022
Initiation Date
June 24, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
436,990

Product Description

The Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin-coated capillary tube) or venipuncture. The sample is applied to an Cholestech LDX¿ cassette. Description/Part: Cholestech LDX Lipid Profile GLU Cassette/10-991, 99021, 14-531; Cholestech LDX TC HDL GLU Cassette/10-990; Cholestech LDX Lipid Profile Cassette/10-989; Cholestech LDX TC HDL Cassette/10-987

Reason for Recall

Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).

Distribution Pattern

US: IL, NV, FL, LA, CA, MA, MN, NC, AK, NE, TX, PA, RI, SC, TN, OK, KY, GA, NY, MO, CO, AZ, OH, HI, VA, KS, IN, AL. OUS: GB, ZA, NO, NL, FI, IE, IQ, IT, CH, DE, ES, HU, SI, NA, PT, IS, NG, PL, CY, TN, BH, ID, MY, AU, SG, JP, IN, BD, KR, HK, BR, CO, AR, CA

Code Information

All lots of Part/UDI-DI: 10-991/00893038002654, 99021/00899722002344, 14-531/04571226470831, 10-990/00893038002647, 10-989/00893038002630, 10-987/00893038002616