Z-2357-2016 Class I Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 31, 2016
- Initiation Date
- July 11, 2016
- Termination Date
- July 20, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
Product Description
Alere INRatio2 PT/INR Professional Testing System Model Number: 0200431 Product Usage: The INRatio 2 PT Monitoring system is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The INRatio 2 PT Monitoring system is intended for use outside the body (in vitro diagnostic use) by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The INRatio 2 PT Monitoring System is not intended to be used for screening purpose.
Reason for Recall
Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
Distribution Pattern
Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Malaysia, Maldives, Netherlands, Russia, Rwanda, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad, Ukraine, Uruguay, and Vietnam.
Code Information
All Lots All Serial Numbers