Z-2359-2016 Class I Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 31, 2016
- Initiation Date
- July 11, 2016
- Termination Date
- July 20, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
Product Description
Alere INRatio2 Replacement Monitor (Home) Model Number: 0200457 Product Usage: Alere INRatio 2 PT/INR Home Monitoring System: The Alere INRatio 2 PT/INR Home Monitoring System, consisting of the INRatio 2 Home Monitor and INRatio PT/INR test strip, is used for quantitative determination of international normalized ratio (INR) in fresh capillary whole blood to monitor the effect of warfarin therapy on clotting time by properly selected suitably trained users (by prescription for home use or other order of a treating physician). Patients must be stabilized (>6 weeks) on warfarin therapy. The Alere INRatio¿2 PT/INR Home Monitoring System is intended for use outside of the body (in vitro diagnostic use). The Alere INRatio¿2 PT/INR Home Monitoring System is not intended to be used for screening purposes.
Reason for Recall
Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
Distribution Pattern
Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Malaysia, Maldives, Netherlands, Russia, Rwanda, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad, Ukraine, Uruguay, and Vietnam.
Code Information
All Lots All Serial Numbers