Z-1514-2022 Class II Ongoing

Recalled by Olympus Corporation Of The Americas — Center Valley, PA

FDA device recall Z-1514-2022 was initiated by Olympus Corporation Of The Americas on Jul 7, 2022 and is designated Class II. Reason for recall: Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility… The recall status is ongoing. Affected quantity: 2 units.

Recall Details

Product Type
Devices
Report Date
Aug 17, 2022
Initiation Date
Jul 7, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product Description

ShockPulse-SE Lithotripsy Generator Model/Serial: SPL-S

Reason for Recall

Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment

Distribution Pattern

US Nationwide distribution in the states of AL, NM.

Code Information

UDI-DI: 00821925044197 Serial numbers: CG5085 and CG6009