Z-1546-2024 Class II Ongoing

Recalled by Beckman Coulter, Inc. — Chaska, MN

FDA device recall Z-1546-2024 was initiated by Beckman Coulter, Inc. on Mar 5, 2024 and is designated Class II. Reason for recall: Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can… The recall status is ongoing. Affected quantity: 17 units.

Recall Details

Product Type
Devices
Report Date
Sep 11, 2024
Initiation Date
Mar 5, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
17 units

Product Description

DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.

Reason for Recall

Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking, leading to delays in reporting patient test results, exposure to physical and chemical hazards, and damage to equipment or the environment.

Distribution Pattern

Worldwide - US Nationwide distribution in the state of New Hampshire and the countries of Belgium, Egypt, Italy, Poland, Spain.

Code Information

UDI/DI 15099590732103, Serial Numbers: 300254, 300256, 300257, 300258, 300259, 300263, 300264, 300266, 300268, 300269, 300270, 300271, 300272, 300273, 300275, 300281, and 300282

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1546-2024
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.