Z-1546-2024 Class II Ongoing
FDA device recall Z-1546-2024 was initiated by Beckman Coulter, Inc. on Mar 5, 2024 and is designated Class II. Reason for recall: Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can… The recall status is ongoing. Affected quantity: 17 units.
Recall Details
- Product Type
- Devices
- Report Date
- Sep 11, 2024
- Initiation Date
- Mar 5, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 17 units
Product Description
DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.
Reason for Recall
Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking, leading to delays in reporting patient test results, exposure to physical and chemical hazards, and damage to equipment or the environment.
Distribution Pattern
Worldwide - US Nationwide distribution in the state of New Hampshire and the countries of Belgium, Egypt, Italy, Poland, Spain.
Code Information
UDI/DI 15099590732103, Serial Numbers: 300254, 300256, 300257, 300258, 300259, 300263, 300264, 300266, 300268, 300269, 300270, 300271, 300272, 300273, 300275, 300281, and 300282