FDA device recall Z-1547-2024 was initiated by Olympus Corporation Of The Americas on Mar 1, 2024 and is designated Class II. Reason for recall: Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance… The recall status is ongoing. Affected quantity: 4507 units.
Recall Details
- Product Type
- Devices
- Report Date
- Sep 18, 2024
- Initiation Date
- Mar 1, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4507 units
Product Description
Olympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to cut tissue using high-frequency current within the upper digestive tract Model: KD-640L
Reason for Recall
Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions for Use, which details critical specifications regarding electrosurgical unit compatibility and output due to reports of serious injury
Distribution Pattern
Nationwide
Code Information
UDI-DI: 04953170208423 All lots numbers