Z-1550-2014 Class II Terminated

Recalled by GE Healthcare — Laurel, MD

FDA device recall Z-1550-2014 was initiated by GE Healthcare on Apr 2, 2014 and is designated Class II. Reason for recall: Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver. The recall status is terminated (terminated May 11, 2016). Affected quantity: 15 units.

Recall Details

Product Type
Devices
Report Date
May 14, 2014
Initiation Date
Apr 2, 2014
Termination Date
May 11, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 units

Product Description

Panda¿ Wall-Mount Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitoring feature used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by a SpO2 sensor).

Reason for Recall

Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.

Distribution Pattern

Worldwide Distribution.

Code Information

HDJS52856 HDJS52857 HDJS52858 HDJS52859 HDJS52860 HDJS52861 HDJS52862 HDJS52863 HDJS52864 HDJS52865 HDJS52800 HDJS52801 HDJS52802 HDJS52803 HDJS52804

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1550-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.