Z-1563-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

FDA device recall Z-1563-2026 was initiated by Medline Industries, LP on Jan 28, 2026 and is designated Class II. Reason for recall: Fabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If… The recall status is ongoing. Affected quantity: 104,004.

Recall details

Product Type
Devices
Report Date
Mar 18, 2026
Initiation Date
Jan 28, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
104,004

Product description

Medline Medcrest Surgical Gowns: MDTGXC4J5XL DBQ-GOWN,SURG,XALT,L4, CRTCL CVR, GRN, 5 MDTGXC4JL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,L MDTGXC4JXL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,XL MDTGXC4JXXL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,XX MDTGXP4JL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,L MDTGXP4JXL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XL MDTGXP4JXXL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XX MDTL1002862EF DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XL MDTZ1000893AJ DBD-GOWN,SRG,PNL CVR, GREEN,SNPS,XL

Reason for recall

Fabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If delaminated product is used and surgical strikethrough occurs, there may be risk of infection.

Distribution pattern

Domestic distribution nationwide US. International distribution to Panama, United Arab Emirates, Virgin Islands, Saudi Arabia, India, Chile, and Canada.

Code information

UDI-DI (each/case) 10197344014002 20197344014009 10888277189188 40888277189189 10884389189191 40884389189192 10884389193297 40884389193298 10888277189218 40888277189219 10884389189221 40884389189222 10193489040111 20193489040118 10884389189221 40884389189222 10193489021820 20193489021827 All lots affected

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1563-2026
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1563-2026. Report a data issue: hello@regdatalab.com.