Z-1575-2023 Class II Ongoing

Recalled by Diagnostica Stago, Inc. — Parsippany, NJ

Recall Details

Product Type
Devices
Report Date
May 17, 2023
Initiation Date
March 22, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 analyzers

Product Description

STA R Max fully automatic clinical instrument designed to perform tests on human plasmas, Reference Numbers 59013 and 59026

Reason for Recall

An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error reported by the analyzer (error 13).

Distribution Pattern

US Nationwide distribution in the states of IL, MA, MN, NE, TX, and WI.

Code Information

UDI-DI: 036074505901; Serial Numbers: CP86010393 DB68093186 CP87010814 CP86070669 CP86090743 CP86090744