Z-1575-2023 Class II Ongoing
FDA device recall Z-1575-2023 was initiated by Diagnostica Stago, Inc. on Mar 22, 2023 and is designated Class II. Reason for recall: An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error… The recall status is ongoing. Affected quantity: 6 analyzers.
Recall details
- Product Type
- Devices
- Report Date
- May 17, 2023
- Initiation Date
- Mar 22, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6 analyzers
Product description
STA R Max fully automatic clinical instrument designed to perform tests on human plasmas, Reference Numbers 59013 and 59026
Reason for recall
An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error reported by the analyzer (error 13).
Distribution pattern
US Nationwide distribution in the states of IL, MA, MN, NE, TX, and WI.
Code information
UDI-DI: 036074505901; Serial Numbers: CP86010393 DB68093186 CP87010814 CP86070669 CP86090743 CP86090744
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1575-2023
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026