Z-1993-2025 Class II Ongoing

Recalled by Diagnostica Stago, Inc. — Parsippany, NJ

Recall Details

Product Type
Devices
Report Date
June 25, 2025
Initiation Date
May 21, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
939 US ( 2,810 total worldwide)

Product Description

STA - Owren Koller. Model/Catalog Number: 00360 STA - Owren-Koller is a buffer solution intended for use as a diluent for reagents and human plasmas in coagulation tests performed with analyzers of the STA-R, STA Compact, STA Satellite and ST art families.

Reason for Recall

Potential for shorter clotting times in clotting tests which have a dilution, leading to possible overestimate of around 1%

Distribution Pattern

The affected lot was distributed to customers in various US states in October and November 2024. The lot was also distributed in other countries including Canada. The lot was provided to two Distributors in US Territories: MedPharm Corp (Guam) and Isla Lab Products (Puerto Rico); MedPharm will be notified by Stago's Asia-Pacific business unit in the coming days and is not part of this report.

Code Information

UDI: (01)13607450003601(11)240331(17)260331(10)270902(241)00360. Lot Number: 270902.