Z-3047-2026 Class II Ongoing

Recalled by Diagnostica Stago, Inc. — Parsippany, NJ

FDA device recall Z-3047-2026 was initiated by Diagnostica Stago, Inc. on July 29, 2026 and is designated Class II. Reason for recall: A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified The recall status is ongoing. Affected quantity: 5220 units.

Recall Details

Product Type
Devices
Report Date
September 9, 2026
Initiation Date
July 29, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5220 units

Product Description

STA Staclot Protein S, Model/Catalog Number: 00746-Reagent Lot number 273074; qualitative and quantitative factor deficiency test

Reason for Recall

A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified

Distribution Pattern

US Nationwide distribution in the states and territory of NJ, NY, NJ, VA, PA, FL, AL, GA, TN, OH, IL, MO, NE, TX, CO, UT, CA, PR.

Code Information

UDI : (01)3607450007466(11)250331(17)270331(10)273074(241)00746, Exp 3/31/2027