Z-1576-2023 Class II Ongoing

Recalled by Diagnostica Stago, Inc. — Parsippany, NJ

Recall Details

Product Type
Devices
Report Date
May 17, 2023
Initiation Date
March 22, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 analyers

Product Description

STA Compact Automated Multi-Parametric Analyzer, Reference Numbers 58602 and 58612

Reason for Recall

An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error reported by the analyzer (error 13).

Distribution Pattern

US Nationwide distribution in the states of IL, MA, MN, NE, TX, and WI.

Code Information

UDI-DI: 036074505860; Serial Numbers: 7096527 7116678