Z-1576-2023 Class II Ongoing

Recalled by Diagnostica Stago, Inc. — Parsippany, NJ

FDA device recall Z-1576-2023 was initiated by Diagnostica Stago, Inc. on Mar 22, 2023 and is designated Class II. Reason for recall: An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error… The recall status is ongoing. Affected quantity: 2 analyers.

Recall details

Product Type
Devices
Report Date
May 17, 2023
Initiation Date
Mar 22, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 analyers

Product description

STA Compact Automated Multi-Parametric Analyzer, Reference Numbers 58602 and 58612

Reason for recall

An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error reported by the analyzer (error 13).

Distribution pattern

US Nationwide distribution in the states of IL, MA, MN, NE, TX, and WI.

Code information

UDI-DI: 036074505860; Serial Numbers: 7096527 7116678

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1576-2023
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Diagnostica Stago, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1576-2023. Report a data issue: hello@regdatalab.com.