Z-1596-2016 Class II Terminated
FDA device recall Z-1596-2016 was initiated by Biomerieux Inc on Feb 24, 2016 and is designated Class II. Reason for recall: Product may lead to false susceptible results of Staphylococcus spp., Enterococcus spp., and S. agalactiae to Minocycline. The recall status is terminated (terminated May 26, 2017). Affected quantity: 337 Kits of 20 Cards/Kit.
Recall Details
- Product Type
- Devices
- Report Date
- May 11, 2016
- Initiation Date
- Feb 24, 2016
- Termination Date
- May 26, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 337 Kits of 20 Cards/Kit
Product Description
VITEK 2 Gram Positive Susceptibility Test Cards (AST-P605), REF 22 325. 20 Cards/Kit. bioMerieux, Inc, Durham, North Carolina 27712. VITEK 2 is an automated system consisting of instruments, software, and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria. The VITEK 2 System is not considered a life supporting/sustaining device and does not deliver any type of energy or substance to the patient. The VITEK 2 Gram Positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.
Reason for Recall
Product may lead to false susceptible results of Staphylococcus spp., Enterococcus spp., and S. agalactiae to Minocycline.
Distribution Pattern
Worldwide Distribution to China Taiwan region; No US/military/government distribution.
Code Information
Lot/Serial: 485368220 Expiry: 12-JAN-17