Z-3231-2026 Class II Ongoing

Recalled by Biomerieux Inc — Hazelwood, MO

FDA device recall Z-3231-2026 was initiated by Biomerieux Inc on Jul 24, 2026 and is designated Class II. Reason for recall: The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false… The recall status is ongoing. Affected quantity: 97680 units.

Recall Details

Product Type
Devices
Report Date
Sep 30, 2026
Initiation Date
Jul 24, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
97680 units

Product Description

VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811

Reason for Recall

The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.

Distribution Pattern

International distribution to the countries of Austria, Germany, Luxembourg.

Code Information

UDI-DI 03573026637712; Lot Number 1333393504

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3231-2026
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.