Z-1609-2013 Class II Terminated

Recalled by Baxter Healthcare Corp. — Round Lake, IL

FDA device recall Z-1609-2013 was initiated by Baxter Healthcare Corp. on Jun 7, 2013 and is designated Class II. Reason for recall: Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the Luer. The recall status is terminated (terminated Jun 4, 2014). Affected quantity: 601,230 total.

Recall Details

Product Type
Devices
Report Date
Jul 3, 2013
Initiation Date
Jun 7, 2013
Termination Date
Jun 4, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
601,230 total

Product Description

Brand Name: Coiled Tube Infusors (Portable Elastomeric Infusion Systems). Baxter's Infusor SV Elastomeric Infusion Device Device Listing: D020871 Baxter's lnfusors SV Elastomeric Infusion Device is a single-use, disposable elastomeric infusion pump indicated tor patients requiring slow, continuous intravenous, intra-arterial, subcutaneous, or epidural administration of medications at a constant flow rate.

Reason for Recall

Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the Luer.

Distribution Pattern

Worldwide Distribution - USA (nationwide), Puerto Rico, and Internationally to Brazil, Colombia, Panama, Mexico, Singapore Japan, Australia, New Zealand, Hong Kong, India, Thailand, Taiwan, and Korea.

Code Information

Product Code: 2C1073KJP Lot Numbers: 08M002, 08N007, 08N029, 08N030, 08N034, 08N069, 08N070, 09A001

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1609-2013