Z-1676-2022 Class II Ongoing

Recalled by Ortho-Clinical Diagnostics, Inc. — Rochester, NY

Recall Details

Product Type
Devices
Report Date
September 14, 2022
Initiation Date
July 19, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1148 units

Product Description

VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461

Reason for Recall

Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer. Potential exists for either a positive or negative shift in results reported by the VITROS System.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda DV04, Brazil, Canada L3R 4G5 , Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.

Code Information

UDI-DI: 10758750031610 Systems start with 7600XXXX where XXXX is unique for each analyzer