Z-1677-2013 Class II Terminated

Recalled by Baxter Healthcare Corp. — Deerfield, IL

FDA device recall Z-1677-2013 was initiated by Baxter Healthcare Corp. on Jul 31, 2009 and is designated Class II. Reason for recall: There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investigation of these reports indicates the incidence of… The recall status is terminated (terminated Mar 8, 2016). Affected quantity: Code 2C1008KP: 56,400 units; Code 2C1009KP: 180,004 units; Code 2C1063KP: 154,937 units; Code 2C1087KP: 43,521 units…

Recall Details

Product Type
Devices
Report Date
Jul 17, 2013
Initiation Date
Jul 31, 2009
Termination Date
Mar 8, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Code 2C1008KP: 56,400 units; Code 2C1009KP: 180,004 units; Code 2C1063KP: 154,937 units; Code 2C1087KP: 43,521 units; Code 2C1156KP: 7,908 units

Product Description

Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices that deliver continuous medication for periods of 30 minutes up to 7 days. They are designed for medication therapies requiring continuous infusion. The infusion pumps can be utilized for slow, continuous delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), and subcutaneous or epidural infusion of medications directly into an intraoperative site or subcutaneously for post-operative pain management.

Reason for Recall

There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investigation of these reports indicates the incidence of leakage has only occurred after pumps are filled, primed, re-capped (using the blue winged Luer cap) and stored for a period of time, typically overnight. In addition, the reported instances of leakage

Distribution Pattern

Worldwide Distribution - USA (nationwide), Puerto Rico and Internationally to Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, Columbia, Czech Republic, Denmark (including Faroe Islands), Ecuador, Finland, France (including Martinique and Reunion Islands), Germany, Greece, Iceland, India, Ireland, Isreal, Italy, Luxemburg, Japan, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweded, Switzerland, Thailand, Turkey, United Arab Emirates, and United Kingdom.

Code Information

*** 1) Product Code: 2C1008KP; Lots: 08M036, 08N041, 09A030, 09B044, 09C053, 09D018, 09E019, 09F005; *** 2) Product Code: 2C1009KP; Lots: 08M016, 08M035, 08M066, 09A010, 09A052, 09A065, 09A080, 09B038, 09B048, 09B060, 09B074, 09C03, 09C016, 09C044, 09C060, 09D011, 09D042, 09D049, 09D054, 09E021, 09E043, 09F002; *** 3) Product Code: 2C1063KP; Lots: 08K078, 08M014, 08M060, 08N022, 08N053, 08N074, 09A032, 09A057, 09B006, 09B014, 09C007, 09C017, 09C054, 09C077, 09D015, 09D041, 09E012,09E018, 09E022, 09E044; *** 4) Product Code: 2C1087KP; Lots: 08K074, 08M050, 08N051, 09A053, 09B063, 09C041, 09D044, 09E073; *** 5) Product Code: 2C1156KP; Lots: 08M063, 09C042

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1677-2013
openFDA file date
Oct 27, 2026
Loaded by RegDataLab
Oct 29, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.