Z-1682-2021 Class II Terminated

Recalled by Pentax of America Inc — Montvale, NJ

Recall Details

Product Type
Devices
Report Date
June 2, 2021
Initiation Date
April 21, 2021
Termination Date
February 5, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
108 units

Product Description

Pentax Medical Video Duodenoscope - Product Usage: intended to be used with endoscopic devices and other ancillary equipment to provide optical visualization of (via a video monitor), and therapeutic access to, Biliary Tract via the Upper Gastrointestinal Tract, Model Number: ED34-i10T2.

Reason for Recall

IFU updated: The warning section of the IFU for both the OE-A63 distal end cap and ED34-i10T2 duodenoscope has been updated to notify users of the associated risks with the distal end cap (OE-A63) unexpectedly becoming detached during a procedure and updated to notify users of what immediate actions should be taken in case the event occurs. This can result in unforseen events such as mucosal injury, lacerations, or bleeding of the patient. Detachment of the distal end cap (OE-A63) into the oral cavity of the patient may also result in aspiration

Distribution Pattern

US Nationwide distribution in the states of Alabama, California, Louisiana, Maryland, Michigan, Missouri, New Jersey, North Carolina, Pennsylvania, Texas, Washington, D.C.

Code Information

All units