Z-0293-2021 Class II Terminated

Recalled by Pentax Of America Inc — Montvale, NJ

FDA device recall Z-0293-2021 was initiated by Pentax Of America Inc on Sep 22, 2020 and is designated Class II. Reason for recall: There is an intermittent software issue that could affect the systems, in which an exam video for one patient (Patient A) might be copied to another patient (Patient B). This does not occur at the… The recall status is terminated (terminated May 16, 2022). Affected quantity: 39 total.

Recall details

Product Type
Devices
Report Date
Nov 4, 2020
Initiation Date
Sep 22, 2020
Termination Date
May 16, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
39 total

Product description

9310HD Digital Video Capture Module with software version 3.4.0 or higher configured with 9263 endoPortal

Reason for recall

There is an intermittent software issue that could affect the systems, in which an exam video for one patient (Patient A) might be copied to another patient (Patient B). This does not occur at the time the user performs and initially reviews an exam; it is not evident until a follow-up review occurs.

Distribution pattern

US Nationwide and Canada

Code information

All systems with version 3.4.0 or higher; Computer Serial Numbers: 84086-01 84533-01 84090-16 84090-15 84090-14 84090-11 84090-09 84090-10 84090-08 84088-13 84088-11 84528-08 84087-06 84087-03 84538-06 84088-14 84089-02 84089-06 84089-03 84529-04 84528-05 81536-06 84091-01 84089-16 81536-12 84532-11 81538-05 84090-02 84089-13 84089-12 84090-06 84087-04 84531-12 84531-11 84531-10 84531-13 84091-16 84533-11 84533-12

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0293-2021
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Pentax Of America Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Pentax Of America Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Pentax Of America Inc →

Source: openFDA record for Z-0293-2021. Report a data issue: hello@regdatalab.com.