Z-1698-2014 Class II Terminated

Recalled by Leica Microsystems, Inc. — Buffalo Grove, IL

FDA device recall Z-1698-2014 was initiated by Leica Microsystems, Inc. on Feb 26, 2014 and is designated Class II. Reason for recall: Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (Product Code: NCL-L-GSTpi-438) does not function as intended up to the expiry date on the product labeling. This was detected… The recall status is terminated (terminated Aug 14, 2017). Affected quantity: 33 units.

Recall Details

Product Type
Devices
Report Date
Jun 11, 2014
Initiation Date
Feb 26, 2014
Termination Date
Aug 14, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
33 units

Product Description

Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (NCL-L-GSTpi-438), a liquid tissue culture supernatant. For in vitro diagnostic use.

Reason for Recall

Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (Product Code: NCL-L-GSTpi-438) does not function as intended up to the expiry date on the product labeling. This was detected through an ongoing stability program. There is a linkage between the age of the product and staining intensity.

Distribution Pattern

Nationwide Distribution including AZ, CA, FL, and NY.

Code Information

Product Code: NCL-L-GSTpi-438; Lot Numbers and Expiration Dates: 6018777, Expires 2014-05; 6023260, Expires 2014-12; 6026297, Expires 2015-05; 6026980, Expires 2015-06

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)