Z-1704-2014 Class II Terminated

Recalled by Hospira Inc. — Lake Forest, IL

FDA device recall Z-1704-2014 was initiated by Hospira Inc. on Jan 24, 2014 and is designated Class II. Reason for recall: There is the potential for the door roller assembly on the Plum LifeCare 5000 Series and Plum XL families of infusers to break. In the event of a broken door roller pin, the door cannot appropriately… The recall status is terminated (terminated Nov 7, 2016). Affected quantity: List Number 11781: 1,828 pumps; List Number 11845: 13,700 pumps; List Number 11846: 22,671 pumps.

Recall Details

Product Type
Devices
Report Date
Jun 11, 2014
Initiation Date
Jan 24, 2014
Termination Date
Nov 7, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
List Number 11781: 1,828 pumps; List Number 11845: 13,700 pumps; List Number 11846: 22,671 pumps

Product Description

Plum XL3 Multi line Infusion System, Micro Macro Plum XL3 Multi line Infusion System, and Micro Macro Plum XL Infusion System. Infusion pumps for intravenous. arterial, short-term epidural, and parenteral administration of general I.V. fluids, medications, nutritional fluids, and blood/blood products to patients in hospital and home care environments.

Reason for Recall

There is the potential for the door roller assembly on the Plum LifeCare 5000 Series and Plum XL families of infusers to break. In the event of a broken door roller pin, the door cannot appropriately lock the cassette in the right position. Depending on the conditions of a broken door roller pin, a number of events can occur. The door cannot be closed or a cassette/door alarm or a false occlusion

Distribution Pattern

Worldwide Distribution - USA nationwide including District of Columbia, Puerto Rico, and US Virgin Islands; Australia, Argentina, Bahrain, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Israel, Italy, Jordan, Kuwait, Luxembourg, Mexico, Oman, Peru, Poland, Portugal, Qatar, Russia, San Marino, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Yemen.

Code Information

List Number: 11781; Serial Numbers: All. List Number: 11845; Serial Numbers: All. List Number: 11846; Serial Numbers: All.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1704-2014
openFDA file date
Oct 27, 2026
Loaded by RegDataLab
Oct 29, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.