Z-1710-2014 Class II Terminated

Recalled by Leica Microsystems, Inc. — Buffalo Grove, IL

FDA device recall Z-1710-2014 was initiated by Leica Microsystems, Inc. on Jan 10, 2014 and is designated Class II. Reason for recall: Product does not function as intended up to the expiry date on the product labeling. There is a link between the age of the product and staining intensity. The recall status is terminated (terminated Sep 5, 2017). Affected quantity: 145 units.

Recall Details

Product Type
Devices
Report Date
Jun 11, 2014
Initiation Date
Jan 10, 2014
Termination Date
Sep 5, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
145 units

Product Description

Liquid Mouse Monoclonal Antibody CD15 (NCL-L-CD15), a liquid tissue culture supernatant containing 15 mM sodium azide as a preservative. For in vitro diagnostic use.

Reason for Recall

Product does not function as intended up to the expiry date on the product labeling. There is a link between the age of the product and staining intensity.

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of CA, CT FL, IL, MO, MS, NC, NY, PA, and the countries of Argentina, Australia, Brazil, Colombia, Italy, Malaysia, New Zealand, Nigeria, Poland, Romania, Russia, Saudi Arabia, Serbia, South Africa, Sweden, Taiwan, Tunisia, Turkey, and United Kingdom

Code Information

Product Code: NCL-L-CD15; Lot Numbers and Expiration Dates: 6017936, Expires 2014-08; 6015073, Expires 2014-03

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1710-2014
openFDA file date
Oct 27, 2026
Loaded by RegDataLab
Oct 29, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.