Z-1710-2022 Class II Ongoing

Recalled by Cook Incorporated — Bloomington, IN

Recall Details

Product Type
Devices
Report Date
September 21, 2022
Initiation Date
July 19, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
585 units

Product Description

Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPW-35-150 ORDER NUMBER (GPN):G56149. For vascular catheter positioning and exchange in diagnostic and interventional procedures.

Reason for Recall

Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.

Distribution Pattern

Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA

Code Information

UDI-DI: (01)00827002561496- Lot Numbers: 14070921, 14108915, 14108916, 14129152, 14141997, 14170230