Z-1935-2026 Class I Ongoing

Recalled by Cook Incorporated — Bloomington, IN

Recall Details

Product Type
Devices
Report Date
May 13, 2026
Initiation Date
April 2, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
271

Product Description

Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip configurations (straight, PIG, and VCF).

Reason for Recall

Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.

Distribution Pattern

US distribution nationwide. OUS distribution to Brunei, Brazil, Canda, China, Colombia, Costa Rica, Hong Kong, Israel, India, Japan, South Korea, Macau, New Zealand, Singapore, El Salvador, Taiwan, Uruguay

Code Information

UDI-DI 00827002473089 Lots 16858459 16869145 16857138 16763324 16750208 16752733 16821053 16713191