Z-1718-2022 Class II Ongoing

Recalled by Cook Incorporated — Bloomington, IN

FDA device recall Z-1718-2022 was initiated by Cook Incorporated on Jul 19, 2022 and is designated Class II. Reason for recall: Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised. The recall status is ongoing. Affected quantity: 15 units.

Recall Details

Product Type
Devices
Report Date
Sep 21, 2022
Initiation Date
Jul 19, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 units

Product Description

Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWS-35-80 ORDER NUMBER (GPN): G56151. For vascular catheter positioning and exchange in diagnostic and interventional procedures.

Reason for Recall

Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.

Distribution Pattern

Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA

Code Information

UDI-DI: (01)00827002561519- Lot Numbers: 14156572