Z-1726-2014 Class II Terminated
FDA device recall Z-1726-2014 was initiated by Leica Microsystems, Inc. on Sep 11, 2013 and is designated Class II. Reason for recall: The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6… The recall status is terminated (terminated Apr 24, 2017). Affected quantity: 103 total.
Recall Details
- Product Type
- Devices
- Report Date
- Jun 18, 2014
- Initiation Date
- Sep 11, 2013
- Termination Date
- Apr 24, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 103 total
Product Description
Leica Ceiling Mount MS-1F & MC-1F for use with the M500 N operating microscope. Article Number 10446657 utilizes a supply voltage of 110 V. Article Number 10446658 utilizes a supply voltage of 220 V. Intended to be used for improving the visibility of objects through magnification and illumination.
Reason for Recall
The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).
Distribution Pattern
Worldwide Distribution -- USA, including the states of AZ, IL, MA, MO, NC, OR, TX, VA, and WI; and, the countries of Australia, Austria, Chile, China, Colombia, Czech Republic, Greece, Israel, Italy, Japan, Malaysia, Netherlands, Norway, Oman, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, and United Kingdom.
Code Information
Leica Ceiling Mounts MS-1F, MC-1F; Article Numbers: 10446657, 10446658