Z-1755-2024 Class II Ongoing

Recalled by Johnson & Johnson Surgical Vision, Inc. — Irvine, CA

Recall Details

Product Type
Devices
Report Date
May 8, 2024
Initiation Date
March 14, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3,428 units

Product Description

Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.

Reason for Recall

Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms.

Distribution Pattern

U.S. Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, LA, MA, MD, ME, MN, MO, NC, NJ, NY, OR, PA, SC, TX, VA, WA, and WI.

Code Information

Part Number: OPOHF21L UDI-DI/GTIN code: 05050474573376 Lot Numbers: 60477723 60479361 60505085 60508179 60508923