Z-1755-2024 Class II Ongoing

Recalled by Johnson & Johnson Surgical Vision, Inc. — Irvine, CA

FDA device recall Z-1755-2024 was initiated by Johnson & Johnson Surgical Vision, Inc. on Mar 14, 2024 and is designated Class II. Reason for recall: Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially… The recall status is ongoing. Affected quantity: 3,428 units.

Recall details

Product Type
Devices
Report Date
May 8, 2024
Initiation Date
Mar 14, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3,428 units

Product description

Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.

Reason for recall

Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms.

Distribution pattern

U.S. Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, LA, MA, MD, ME, MN, MO, NC, NJ, NY, OR, PA, SC, TX, VA, WA, and WI.

Code information

Part Number: OPOHF21L UDI-DI/GTIN code: 05050474573376 Lot Numbers: 60477723 60479361 60505085 60508179 60508923

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1755-2024
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Johnson & Johnson Surgical Vision, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1755-2024. Report a data issue: hello@regdatalab.com.