Z-0735-2024 Class II Ongoing
FDA device recall Z-0735-2024 was initiated by Johnson & Johnson Surgical Vision, Inc. on Dec 12, 2023 and is designated Class II. Reason for recall: Due to intraocular lens containing an angle out of specification (OOS). The cylinder axis marks do not meet product specifications for the allowable angle between Low Power Meridian (LPM) and toric… The recall status is ongoing. Affected quantity: 36.
Recall details
- Product Type
- Devices
- Report Date
- Jan 24, 2024
- Initiation Date
- Dec 12, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 36
Product description
Reason for recall
Distribution pattern
International distribution to the country of Japan.