Z-1006-2023 Class II Ongoing

Recalled by Johnson & Johnson Surgical Vision, Inc. — Irvine, CA

FDA device recall Z-1006-2023 was initiated by Johnson & Johnson Surgical Vision, Inc. on Dec 7, 2022 and is designated Class II. Reason for recall: Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the… The recall status is ongoing. Affected quantity: 7644.

Recall details

Product Type
Devices
Report Date
Feb 1, 2023
Initiation Date
Dec 7, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7644

Product description

FUSION Dual Pump Packs, REF: OPO73

Reason for recall

Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

Distribution pattern

US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom

Code information

UDI-DI/Lots: 15050474602080/60373238, 60394037, 60308197

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1006-2023
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Johnson & Johnson Surgical Vision, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1006-2023. Report a data issue: hello@regdatalab.com.