Z-3064-2026 Class II Ongoing
FDA device recall Z-3064-2026 was initiated by Johnson & Johnson Surgical Vision, Inc. on July 31, 2026 and is designated Class II. Reason for recall: Surgical instructions sent for viscoelastic products used during ophthalmic viscosurgical cataract procedures including in part: If resistance is encountered during insertion, withdraw cannula and rea… The recall status is ongoing. Affected quantity: 111,478.
Recall Details
- Product Type
- Devices
- Report Date
- September 9, 2026
- Initiation Date
- July 31, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 111,478
Product Description
HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU); HEALON DUET PRO II, REF: 10301016 (EU)
Reason for Recall
Surgical instructions sent for viscoelastic products used during ophthalmic viscosurgical cataract procedures including in part: If resistance is encountered during insertion, withdraw cannula and reassess the insertion conditions (i.e. insertion angle, incision size, etc.) before proceeding because forcing the cannula may cause corneal abrasion, iris trauma or descemet's membrane detachment.
Distribution Pattern
Worldwide - US Nationwide distribution including in the states of PA, AL, AZ, AR, CA, FL, IL, IN, IA, KY, LA, ME, MI, MN, MS, NJ, NY, NC, OH, OR, TN, TX, WA, WV, HI, CO, MD, VA, WI, ID, NH, SD, OK, MA, NM, DE, GA, SC, CT, NV, MO, RI, KS, ND and the countries of CA, BE, CH, DE, DK, ES, FR GB, HU, IS, IT, NL, NO, PT, SE.
Code Information
REF/UDI-DI/Lot: 10300012/5050474755093/UA32232, UA32250, UA32268, UA32290, UA32292, UA32303, UA32371, UP31203, UP31205; 10300016/5050474755116/UA32447, UA32448, UA32449, UB33008, UB33026, UB33082, UB33083, UB33175; 10301012/5050474755109/UB33014, UB33041, UB33057, UB33060, UB33100, UB33190; 10301016/5050474755123/UB33084, UB33194