Z-1770-2022 Class II Ongoing

Recalled by Fresenius Medical Care Holdings, Inc. — Waltham, MA

FDA device recall Z-1770-2022 was initiated by Fresenius Medical Care Holdings, Inc. on Jun 21, 2022 and is designated Class II. Reason for recall: Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured The recall status is ongoing. Affected quantity: 6 units.

Recall Details

Product Type
Devices
Report Date
Sep 28, 2022
Initiation Date
Jun 21, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 units

Product Description

AquaC UNO H [120V, 60Hz, 1700VA, US-Version- a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-0970-1 Vivonic Part Number: G02000966-US

Reason for Recall

Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured

Distribution Pattern

US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV, NY, PA, TX.

Code Information

UDI: (01)4251285200234(11)200422(21)0CHS3987 (01)4251285200234(11)200526(21)0CHS4108 (01)4251285200234(11)210623(21)1CHS5372 (01)4251285200234(11)210707(21)1CHS5405 (01)4251285200234(11)190305(21)9CHS2881 (01)4251285200234(11)190529(21)9CHS3090 Serial Numbers: 0CHS3987 0CHS4108 1CHS5372 1CHS5405 9CHS2881 9CHS3090

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1770-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.