Z-1794-2013 Class II Terminated

Recalled by Hill-Rom, Inc. — Batesville, IN

FDA device recall Z-1794-2013 was initiated by Hill-Rom, Inc. on May 30, 2013 and is designated Class II. Reason for recall: When the bed is put into the chair egress position, the magnets on the mattress may not hold the bottom of the mattress flat to the bed frame, and the mattress foot section will extend away from the… The recall status is terminated (terminated Apr 17, 2014). Affected quantity: 4,121 Units.

Recall Details

Product Type
Devices
Report Date
Jul 31, 2013
Initiation Date
May 30, 2013
Termination Date
Apr 17, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4,121 Units

Product Description

TotalCare Bed used with AccuMax Quantum VPC mattress AccuMax Quantum Complete mattress, and AccuMax Quantum Convertible mattress The TotalCare Bed System is intended to provide a patient support ideally suited to be used in health care environments. The TotalCare Bed System may be used in a variety of settings including, but not limited to, acute care, including critical care, step down/progressive care, medical/surgical, high acuity sub-acute care, post anesthesia care unit (PACU), and sections of the emergency department (ED). The TotalCare Bed System is capable of being used with a broad patient population as determined appropriate by the caregiver or institution.

Reason for Recall

When the bed is put into the chair egress position, the magnets on the mattress may not hold the bottom of the mattress flat to the bed frame, and the mattress foot section will extend away from the bed at an angle. A patient attempting to get back into the bed may miss the seat and sit on the foot section instead. The patient can slide down to the floor, or fall, with possible injury.

Distribution Pattern

Worldwide distribution: USA (nationwide) and countries of: Canada, Colombia, Nigeria, Kuwait and Saudi Arabia.

Code Information

Bed model numbers: PAU001010784, PAU005010784, P1900, PAK169N10184, PAK169N10384, PA8171210184, PA8175210184, PA8175210384, PA8171210384, PA8171310384, and PA8173210384. The AccuMax Quantum VPC, Complete and Convertible mattresses are not serialized, but were distributed between September 2005 and April 2013

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1794-2013
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.