Z-1813-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 6, 2020
- Initiation Date
- March 18, 2020
- Termination Date
- February 9, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 78,521 units total
Product Description
T12. Incorporates wireless electrocardiographic technology to achieve the real-time acquisition and RF transmission of simultaneous 12-lead ECG data with diagnostic quality while allowing the patient to be ambulatory.
Reason for Recall
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Distribution Pattern
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
Code Information
Model Numbers: T12-XXX-XXXXX, T12S-XXX-XXXXX.