Z-1811-2020 Class II Terminated

Recalled by WELCH ALLYN, INC/MORTARA — Milwaukee, WI

Recall Details

Product Type
Devices
Report Date
May 6, 2020
Initiation Date
March 18, 2020
Termination Date
February 9, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
78,521 units total

Product Description

ELI PC. The ELI PC Service automatically converts resting ECG recordings collected with WAM PC patient cables and WebUpload into DICOM 12-lead ECG Waveform files.

Reason for Recall

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Distribution Pattern

US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.

Code Information

Model Numbers: 11066-XXX-50, 41000-029-XX.