Z-1811-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 6, 2020
- Initiation Date
- March 18, 2020
- Termination Date
- February 9, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 78,521 units total
Product Description
ELI PC. The ELI PC Service automatically converts resting ECG recordings collected with WAM PC patient cables and WebUpload into DICOM 12-lead ECG Waveform files.
Reason for Recall
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Distribution Pattern
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
Code Information
Model Numbers: 11066-XXX-50, 41000-029-XX.