Z-1815-2020 Class II Terminated

Recalled by WELCH ALLYN, INC/MORTARA — Milwaukee, WI

Recall Details

Product Type
Devices
Report Date
May 6, 2020
Initiation Date
March 18, 2020
Termination Date
February 9, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
78,521 units total

Product Description

S4 and patient cable for the S4 Wireless Telemeter with the following device names: CABLE 10 WIRE LF ECG SNAP ENDS AHA GRAY, CABLE 10 WIRE LF ECG SNAP ENDS IEC GRAY, CBL 10WIRE LF LRG ECG SNAP ENDS AHA GRAY, CBL 10WIRE LF LRG ECG SNAP ENDS IEC GRAY. The S4 mobile monitor provides a means to acquire and transmit simultaneous ECG and SpO2 data to a Surveyor Central Station monitoring system while allowing the patient to be ambulatory within the range of the WiFi network.

Reason for Recall

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Distribution Pattern

US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.

Code Information

Model Numbers: S4-Q-ASX-XXX, S4-Q-AXX-XXX, S4-P-A, 9293-061-50, 9293-061-51, 9293-061-52, 9293-061-53.