Z-1825-2022 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- October 5, 2022
- Initiation Date
- August 16, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 488 US, 5 OUS
Product Description
Bentson Plus Wire Guide (Extra Body/20cm Distal Flexibility with Softened Tip), Guidewire; Reference Part Number: TSFBP-35-180 (GPN G09782) TSFBP-35-145 (GPN G09735)
Reason for Recall
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
Distribution Pattern
Domestic distribution nationwide. Foreign distribution worldwide.
Code Information
TSFBP-35-180 (GPN G09782) UDI-DI 00827002097827, Lots 14797763 14797770 14805076 14805079 14813709 14813711; TSFBP-35-145 (GPN G09735) UDI-DI 00827002097353, Lots 14807595 14812037 14812038