Z-1826-2022 Class II Ongoing

Recalled by Cook Incorporated — Bloomington, IN

Recall Details

Product Type
Devices
Report Date
October 5, 2022
Initiation Date
August 16, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
43 OUS

Product Description

Bentson Wire Guide, Guidewire (OUS ONLY); Reference Part Number: TSMB-35-145 (GPN G27041)

Reason for Recall

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Distribution Pattern

Domestic distribution nationwide. Foreign distribution worldwide.

Code Information

TSMB-35-145 (GPN G27041) UDI-DI 00827002270411, Lot 14797970, Expiration Date 16-06-2027