Z-1836-2022 Class II Ongoing

Recalled by Cook Incorporated — Bloomington, IN

Recall Details

Product Type
Devices
Report Date
October 5, 2022
Initiation Date
August 16, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
42 US

Product Description

Newton Wire Guide (LT Taper/10cm Distal Flexibility), Guidewire Reference Part Number/GPN: TSFNA-35-145 G00701 TSCFNA-35-145-3 G00561

Reason for Recall

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Distribution Pattern

Domestic distribution nationwide. Foreign distribution worldwide.

Code Information

TSFNA-35-145 G00701 UDI-DI: 00827002007017 Lots/Expiration Date: 14798674 16-06-2027 TSCFNA-35-145-3 G00561 UDI-DI: 00827002005617 Lots/Expiration Date: 14835600 11-07-2027