Z-1901-2018 Class II Terminated

Recalled by LivaNova USA — Arvada, CO

FDA device recall Z-1901-2018 was initiated by LivaNova USA on Apr 20, 2018 and is designated Class II. Reason for recall: Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain… The recall status is terminated (terminated Feb 5, 2025). Affected quantity: 2408.

Recall Details

Product Type
Devices
Report Date
May 30, 2018
Initiation Date
Apr 20, 2018
Termination Date
Feb 5, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2408

Product Description

Sorin Stockert Heater-Cooler System 3T, a) Item #: 16-02-81, 240 v/60 Hz; b) Item #: 16-02-82, 208 v/60 Hz; c) Item #: 16-02-85, 120 v/60 Hz

Reason for Recall

Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm.

Distribution Pattern

MN, TX, WA, OR, IL, OH, SC, AK, PA, CA, WI, GA, NJ, FL, AZ, KY, MS, MO, LA, AR, MA, MI, MT, CO, TN, NE, NC, VA, NV, HI, PR, WY, AL, OK, IN, CT, ME, IA, SC, KS, NM, SD, DC, MD, ND,

Code Information

Serial #: a) 16S10743 - 6S11708 b) 16S10772 - 16S16840 C) 16S10958 - 16S16847

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1901-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.