Z-1903-2012 Class II Terminated

Recalled by Polymer Technology Systems, Inc. — Indianapolis, IN

FDA device recall Z-1903-2012 was initiated by Polymer Technology Systems, Inc. on May 15, 2012 and is designated Class II. Reason for recall: During a routine product investigation, a creatinine sample with a value less than 1.0 mg/dl was tested on a version 2.60 CardioCek PA meter and the meter provided a result of 0.08mg/dl, which is not… The recall status is terminated (terminated Jul 20, 2012). Affected quantity: 1650 vials.

Recall details

Product Type
Devices
Report Date
Jul 4, 2012
Initiation Date
May 15, 2012
Termination Date
Jul 20, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1650 vials

Product description

PTS Panels Creatinine test strips. PTS Panels Creatinine Test Strips are intended to measure creatinine in whole blood. Creatinine measurements are used in the diagnosis and treatment of renal (kidney) diseases and in the monitoring of renal dialysis. This system is intended for the professional use for accurately and reproducibly measuring creatinine in human whole blood obtained from a finger stick sample

Reason for recall

During a routine product investigation, a creatinine sample with a value less than 1.0 mg/dl was tested on a version 2.60 CardioCek PA meter and the meter provided a result of 0.08mg/dl, which is not the reportable measuring range for the CardioChek PA meter for the creatinine assay.

Distribution pattern

Worldwide Distribution -- Nationwide Distribution including the states of CA and VA., and the countries of China, Poland, Ireland and France.

Code information

Cat No. 1720; and Lot number F103

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1903-2012
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Polymer Technology Systems, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1903-2012. Report a data issue: hello@regdatalab.com.