Z-1963-2014 Class II Terminated

Recalled by Baxter Healthcare Corp. — Deerfield, IL

FDA device recall Z-1963-2014 was initiated by Baxter Healthcare Corp. on Jun 10, 2014 and is designated Class II. Reason for recall: If the universal ingredient (UI) in an active configuration is changed using the Configuration Editor, a flush of the outlet pump tube will not be initiated by the software. It could result in the… The recall status is terminated (terminated Feb 22, 2017). Affected quantity: 1758.

Recall Details

Product Type
Devices
Report Date
Jul 9, 2014
Initiation Date
Jun 10, 2014
Termination Date
Feb 22, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1758

Product Description

EXACTAMIX 1200 Compounder, model numbers 1200-DY, 1200-DX, 1200-DYR and 1200-DXR; EM1200 DY Display, Replacement EM 1200 DY Display. EXACTAMIX 2400 Compounder model numbers 2400-DY, 2400-DX, 2400-DYR and 2400-DYX. Baxa Corporation. An automated pumping system that compounds multiple sterile ingredients into a finished solution in a single patient IV administration bag.

Reason for Recall

If the universal ingredient (UI) in an active configuration is changed using the Configuration Editor, a flush of the outlet pump tube will not be initiated by the software. It could result in the original UI remaining in the tube which could be delivered into the next bag.

Distribution Pattern

US Nationwide

Code Information

model numbers 1200-DY, 1200-DX, 1200-DYR and 1200-DXR; EXACTAMIX 2400 Compounder model numbers 2400-DY, 2400-DX, 2400-DYR and 2400-DYX

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1963-2014
openFDA file date
Oct 27, 2026
Loaded by RegDataLab
Oct 29, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.