Z-1967-2023 Class II

Recalled by Vyaire Medical — Mettawa, IL

Recall Details

Product Type
Devices
Report Date
June 21, 2023
Initiation Date
May 5, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
392 units

Product Description

bellavista 1000e Ventilator, Catalogue Number 301.100.130; Continuous Use Ventilator

Reason for Recall

Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang). When this occurs, if the user continues to interact with the touchscreen, a message will appear that states DeviceSoftware.Application is not responding or a decommission screen will appear on the user interface. The ventilator will issue both an audible and visual alarm. Ventilation continues without interruption with the settings applied prior to the APP Hang error. 2. The potential for a use error has been identified if the operator applies the proposed settings without confirming the settings are suitable for the patient.

Distribution Pattern

US Nationwide Distribution

Code Information

UDI/DI 07640149381115 and 07640149388879, ALL serial numbers with Software Version 6.1.0.2 and 6.1.1