Z-2027-2014 Class II Terminated
FDA device recall Z-2027-2014 was initiated by Hospira Inc. on Jun 6, 2014 and is designated Class II. Reason for recall: Hospira MedNet 5.5, 5.8.1, and 5.8.2 contains software defect where the dosing units of "nanog/kg/min" and "milliUnits/min" are not sent to a Plum A+ device (Version 13.40, 13.41, and 13.4.2) via Auto… The recall status is terminated (terminated Dec 6, 2016). Affected quantity: 13 units.
Recall Details
- Product Type
- Devices
- Report Date
- Jul 16, 2014
- Initiation Date
- Jun 6, 2014
- Termination Date
- Dec 6, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 13 units
Product Description
The Hospira MedNet Medication Management Suite (MMS). Product Usage: The Hospira MedNet Medication Management Suite (MMS) is intended to facilitate networked communication between MMS compatible computer systems and Hospira infusion pumps. The MMS provides trained healthcare professionals with the capability to send, receive, report, and store information from interfaced external systems, and to configure and edit infusion programming parameters. The MMS is intended to provide a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data. All data entry and validation of infusion parameters is performed by a trained healthcare professional according to physicians orders.
Reason for Recall
Hospira MedNet 5.5, 5.8.1, and 5.8.2 contains software defect where the dosing units of "nanog/kg/min" and "milliUnits/min" are not sent to a Plum A+ device (Version 13.40, 13.41, and 13.4.2) via Auto Programming when those dosing units are not properly established in the customized drug library. A dosing unit mismatch of "nanog/kg/min" or "milliUnits/min" on an unmatched medication will produce a
Distribution Pattern
US Nationwide in the states of FL, LA, PA, WA, WY and the country of Hong Kong
Code Information
1) Software Version 5.5, List Number: 16037-04-13; 2) Software Version 5.8.1, List Number: 16037-04-31; 3) Software Version 5.8.2, List Number: 16037-04-33
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2027-2014
- openFDA file date
- Oct 27, 2026
- Loaded by RegDataLab