Z-2029-2024 Class II Ongoing

Recalled by Olympus Corporation Of The Americas — Center Valley, PA

FDA device recall Z-2029-2024 was initiated by Olympus Corporation Of The Americas on Apr 18, 2024 and is designated Class II. Reason for recall: The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image. The recall status is ongoing. Affected quantity: 1 unit.

Recall Details

Product Type
Devices
Report Date
Jun 12, 2024
Initiation Date
Apr 18, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product Description

OES 4000 Hysteroscope, Model/Catalog Number: A4674A

Reason for Recall

The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.

Distribution Pattern

US Nationwide distribution in the state of NY.

Code Information

Model/Catalog Number: A4674A, UDI-DI: 04042761006361, Serial Number: 805904;