Z-2040-2024 Class II Ongoing

Recalled by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING — Tirat Carmel, N/A

FDA device recall Z-2040-2024 was initiated by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING on Apr 15, 2024 and is designated Class II. Reason for recall: There is a potential issue regarding the service handles for the internal lead (Pb) covers in that the handles could loosen or release which could cause a cover to drop potentially resulting in injury… The recall status is ongoing. Affected quantity: 133 devices.

Recall Details

Product Type
Devices
Report Date
Jun 19, 2024
Initiation Date
Apr 15, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
133 devices

Product Description

MyoSPECT System, Model H3912AA and H3912B, and MyoSPECT ES System, Model H3912AB and H3912BD, dedicated cardiac SPECT (single photon emission computerized tomography system).

Reason for Recall

There is a potential issue regarding the service handles for the internal lead (Pb) covers in that the handles could loosen or release which could cause a cover to drop potentially resulting in injury to service personnel due to the weight of the cover.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, LA, MA, MD, MI, MO, NC, ND, NV, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Belgium, Brazil, Canada, Czech Republic, Finland, France, India, Israel, Italy, Japan, Malaysia, Malta, Poland, Singapore, Spain, Taiwan, Turkey, and United Kingdom.

Code Information

MyoSPECT manufactured on or before 2024-02-13: Model H3912AA, GTIN 00195278421586 - Serial numbers NGCC80094, NGCB80033, NGCB80044, NGCB80045, NGCA80017, NGCB80039, NGCA80006, NGCA80011, NGCA80003, NGCA80012, NGCB80052, NGCB80060, NGCC80074, NGCC80081, NGCB80038, NGCD80107, NGCB80070, NGCC80093, NGCD80106, NGCB80026, NGCB80050, NGCB80058, NGCC80084, NGCB80027, NGCB80072, NGCB80032, NGCB80031, NGCC80080, NGCB80067, NGCA80021, NGCA80010, NGCA80015, NGCB80043, NGCC80090, NGCC80091, NGCD80104, NGCB80028, NGCB80022, NGCB80023, NGCB80048, NGCB80035, NGCD80103, NGCC80100, NGCC80083, NGCB80042, NGCC80075, NGCB80068, NGCB80040, NGCB80051, NGCB80064, NGCC80078, NGCC80079, NGCC80087, NGCC80092, NGCD80105, NGCD80108, NGCB80059, NGCA80009, NGCC80086, NGCC80095, NGCA80020, NGCC80088, NGCB80055, NGCB80029, NGCB80041, NGCC80076, NGCA80016, NGCB80036, NGCC80098, NGCB80037, NGCB80046, NGCA80007, NGCC80082, NGCB80061, NGCA80002, NGCC80102, NGCB80057, NGCC80089, NGCB80030, NGCC80097, NGCA80005, NGCB80049, NGCC80101, NGCB80071, NGCA80001, NGCB80066, NGCB80065, NGCC80085, NGCC80077, NGCB80069, NGCA80014, MYGA10001, NGCC80096, NGCB80024, NGCB80047, NGCC80099, NGCB80062, NGCB80034, NGCA80008, NGCA80013, NGCB80025, NGCA80004, NGCA80019, NGCB80056, NGCB80063, NGCB80054, and NGCB80053. Model H3912BC, GTIN 00195278488619 - Serial number MYOC93001. MyoSPECT ES manufactured on or before 2024-02-15: Model H3912AB, GTIN 00195278421579 - Serial numbers NGEB81002, NGED81025, NGEB81007, NGEB81006, NGEC81019, NGEB81012, NGEB81008, NGEC81014, NGEB81005, NGEC81017, NGEC81022, NGEC81021, NGEC81020, NGEB81011, NGEB81010, NGEC81023, NGEC81024, NGEC81013, NGEB81009, NGEB81003, NGEB81004, NGEC81018, NGEC81015, and NGEC81016. Model H3912BD, GTIN 00195278488626 - Serial number MYEB94001.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2040-2024
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.