Z-2065-2019 Class I Terminated

Recalled by Fenwal Inc — Lake Zurich, IL

Recall Details

Product Type
Devices
Report Date
August 21, 2019
Initiation Date
June 24, 2019
Termination Date
March 6, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
83 units

Product Description

Vigilant Agilia, Vigilant Drug'Lib, REF Z073476, versions 1.0 and 1.1 Product Usage: Vigilant Drug Lib is Dose Error Reduction Software that, when used in combination with Volumat MC Agilia infusion pumps, is intended to reduce drug administration errors at the bedside. Vigilant Drug Lib Agilia is a software application running on a computer and intended to configure and generate multiple device configurations, drug libraries, care areas and data sets that can then be uploaded to the Volumat MC Agilia infusion pumps. Intended patient population: The Vigilant Drug Lib software must only be used to create drug libraries for patients whose weight is 250g to 250 kg (0.6 lbs. to 551 lbs.).

Reason for Recall

The firm is correcting four software anomalies and KVO (keep vein open) end of infusion alarm priority in Software versions 1.7 and 1.9a of Volumat MC Agilia Software and versions 1.0 and 1.1 of Vigilant DrugLib

Distribution Pattern

US Nationwide Distribution

Code Information

All serial numbers distributed in the US