Z-2065-2019 Class I Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 21, 2019
- Initiation Date
- June 24, 2019
- Termination Date
- March 6, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 83 units
Product Description
Vigilant Agilia, Vigilant Drug'Lib, REF Z073476, versions 1.0 and 1.1 Product Usage: Vigilant Drug Lib is Dose Error Reduction Software that, when used in combination with Volumat MC Agilia infusion pumps, is intended to reduce drug administration errors at the bedside. Vigilant Drug Lib Agilia is a software application running on a computer and intended to configure and generate multiple device configurations, drug libraries, care areas and data sets that can then be uploaded to the Volumat MC Agilia infusion pumps. Intended patient population: The Vigilant Drug Lib software must only be used to create drug libraries for patients whose weight is 250g to 250 kg (0.6 lbs. to 551 lbs.).
Reason for Recall
The firm is correcting four software anomalies and KVO (keep vein open) end of infusion alarm priority in Software versions 1.7 and 1.9a of Volumat MC Agilia Software and versions 1.0 and 1.1 of Vigilant DrugLib
Distribution Pattern
US Nationwide Distribution
Code Information
All serial numbers distributed in the US