Z-1209-2019 Class II Terminated

Recalled by Fenwal Inc — Lake Zurich, IL

FDA device recall Z-1209-2019 was initiated by Fenwal Inc on Mar 5, 2019 and is designated Class II. Reason for recall: Based on internal investigations, Fresenius Kabi has identified a multifaceted set of events involving the CATSmart Systems, including the device, software, and disposable which may result in the… The recall status is terminated (terminated Oct 5, 2020). Affected quantity: 342 devices.

Recall details

Product Type
Devices
Report Date
May 1, 2019
Initiation Date
Mar 5, 2019
Termination Date
Oct 5, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
342 devices

Product description

Fresenius Kabi CATSmart device (Continuous autotransfusion System). Labeled as the following kits: 1. AT-1 sets and kits (9005104 AT-1 Autotransfusion set, 9108504 ATF 120 Fast Start Kit, and 9108494 ATF 40 Fast Start Kits); 2. AT-3 Autotransfusion sets and kits (Part 9005444 AT-3 set, 9104444 ATF-120 Fast Start Kits (AT-3), or 9108444 ATF-40 Fast Start Kits AT-3) Product Usage: The Fresenius Kabi CATSmart device (Continuous auto transfusion System) is indicated for the processing of autologous shed blood collected interaoperatively and postoperatively to obtain washed packed red blood cells for reinfusion. The CATSmart Auto transfusion System is designed to reduce the need for allogeneic blood transfusion by re-infusing a patient s own lost blood during surgery. The set is used as a disposable component of the autotransfusion system during surgery and/or treatment of trauma to remove waste and improve the safety of autologous re-infused blood (auto transfusion). This is a single-use device.

Reason for recall

Based on internal investigations, Fresenius Kabi has identified a multifaceted set of events involving the CATSmart Systems, including the device, software, and disposable which may result in the centrifuge tubing line making contact with the washing chamber. The contact may result in excessive wear of the centrifuge tubing or in extreme cases a puncture of the tubing.

Distribution pattern

Worldwide distribution - US nationwide in the states of AL, AZ, CA, CO, FL, GA, HI, ID, IL, IN, LA, MA, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, and PR, and country of Canada.

Code information

Product Code: 9005901; Lot Numbers beginning with GM or H. Serial Numbers: 6CTA0209, 6CTA0210, 6CTA0211, 6CTA0212, 6CTA0213, 6CTA0214, 6CTA0215, 6CTA0216, 6CTA0217, 6CTA0218, 6CTA0219, 6CTA0221, 6CTA0222, 6CTA0223, 6CTA0224, 6CTA0225, 6CTA0226, 6CTA0227, 6CTA0228, 6CTA0260, 6CTA0261, 6CTA0262, 6CTA0263, 6CTA0264, 6CTA0265, 6CTA0266, 6CTA0267, 6CTA0268, 6CTA0269, 6CTA0300, 6CTA0301, 6CTA0302, 6CTA0303, 6CTA0304, 6CTA0305, 6CTA0306, 6CTA0307, 6CTA0308, 6CTA0309, 7CTA0328, 7CTA0329, 7CTA0330, 7CTA0331, 7CTA0332, 7CTA0334, 7CTA0335, 7CTA0336, 7CTA0337, 7CTA0358, 7CTA0359, 7CTA0360, 7CTA0361, 7CTA0362, 7CTA0363, 7CTA0364, 7CTA0365, 7CTA0366, 7CTA0367, 7CTA0368, 7CTA0369, 7CTA0370, 7CTA0371, 7CTA0372, 7CTA0373, 7CTA0374, 7CTA0375, 7CTA0376, 7CTA0377, 7CTA0378, 7CTA0379, 7CTA0380, 7CTA0381, 7CTA0392, 7CTA0393, 7CTA0394, 7CTA0395, 7CTA0396, 7CTA0397, 7CTA0398, 7CTA0399, 7CTA0401, 7CTA0402, 7CTA0403, 7CTA0404, 7CTA0405, 7CTA0406, 7CTA0409, 7CTA0410, 7CTA0411, 7CTA0412, 7CTA0413, 7CTA0414, 7CTA0433, 7CTA0434, 7CTA0435, 7CTA0436, 7CTA0437, 7CTA0438, 7CTA0439, 7CTA0440, 7CTA0472, 7CTA0473, 7CTA0474, 7CTA0475, 7CTA0476, 7CTA0477, 7CTA0478, 7CTA0479, 7CTA0480, 7CTA0481, 7CTA0482, 7CTA0483, 7CTA0484, 7CTA0485, 7CTA0486, 7CTA0487, 7CTA0488, 7CTA0489, 7CTA0509, 7CTA0510, 7CTA0511, 7CTA0512, 7CTA0513, 7CTA0514, 7CTA0515, 7CTA0516, 7CTA0517, 7CTA0518, 7CTA0519, 7CTA0520, 7CTA0521, 7CTA0522, 7CTA0523, 7CTA0548, 7CTA0549, 7CTA0550, 7CTA0551, 7CTA0552, 7CTA0553, 7CTA0554, 7CTA0556, 7CTA0557, 7CTA0558, 7CTA0580, 7CTA0581, 7CTA0582, 7CTA0583, 7CTA0584, 7CTA0585, 7CTA0586, 7CTA0587, 7CTA0588, 7CTA0589, 7CTA0590, 7CTA0591, 7CTA0647, 7CTA0648, 7CTA0649, 7CTA0650, 7CTA0651, 7CTA0652, 7CTA0653, 7CTA0654, 7CTA0655, 7CTA0656, 7CTA0657, 7CTA0658, 7CTA0659, 7CTA0660, 7CTA0661, 7CTA0662, 7CTA0663, 7CTA0664, 7CTA0665, 7CTA0666, 7CTA0687, 7CTA0688, 7CTA0689, 7CTA0690, 7CTA0691, 7CTA0692, 7CTA0696, 7CTA0697, 7CTA0698, 7CTA0699, 7CTA0700, 7CTA0701, 7CTA0702, 7CTA0715, 7CTA0716, 7CTA0717, 7CTA0719, 7CTA0720, 7CTA0721, 7CTA0722, 7CTA0723, 7CTA0724, 7CTA0735, 7CTA0736, 7CTA0737, 7CTA0738, 7CTA0739, 7CTA0740, 8CTA0741, 8CTA0742, 8CTA0748, 8CTA0749, 8CTA0754, 8CTA0911, 8CTA0912, 8CTA0913, 8CTA0914, 8CTA0915, 8CTA0916, 8CTA0917, 8CTA0918, 8CTA0919, 8CTA0920, 8CTA0921, 8CTA0922, 8CTA0923, 8CTA0924, 8CTA0925, 8CTA0926, 8CTA0927, 8CTA0928, 8CTA0929, 8CTA0930, 8CTA0931, 8CTA0932, 8CTA0933, 8CTA0934, 8CTA0935, 8CTA0936, 8CTA0937, 8CTA0938, 8CTA0939, 8CTA0940, 8CTA0942, 8CTA0943, 8CTA0944, 8CTA0945, 8CTA0946, 8CTA0947, 8CTA0948, 8CTA0949, 8CTA0950, 8CTA0951, 8CTA0952, 8CTA0953, 8CTA0954, 8CTA0955, 8CTA0956, 8CTA0957, 8CTA0958, 8CTA0959, 8CTA0960, 8CTA0961, 8CTA0962, 8CTA0963, 8CTA0964, 8CTA0965, 8CTA0966, 8CTA0967, 8CTA0968, 8CTA0969, 8CTA0970, 8CTA0971, 8CTA0972, 8CTA0973, 8CTA0974, 8CTA0975, 8CTA0976, 8CTA0977, 8CTA0978, 8CTA0979, 8CTA0980, 8CTA0981, 8CTA0982, 8CTA0983, 8CTA0984, 8CTA0985, 8CTA0986, 8CTA0987, 8CTA0988, 8CTA0989, 8CTA0990, 8CTA0991, 8CTA0992, 8CTA0993, 8CTA0994, 8CTA0996, 8CTA0997, 8CTA0998, 8CTA0999, 8CTA1000, 8CTA1001, 8CTA1002, 8CTA1003, 8CTA1004, 8CTA1005, 8CTA1006, 8CTA1007, 8CTA1008, 8CTA1009, 8CTA1010, 8CTA1011, 8CTA1032, 8CTA1033, 8CTA1037, 8CTA1039, 8CTA1040, 8CTA1041, 8CTA1042, 8CTA1053, 8CTA1057, 8CTA1056, 8CTA1055, 8CTA1054, 8CTA1051, 6CTA0220, 8CTA1050, 8CTA1049, 8CTA1048, 8CTA1047, 8CTA1046, 8CTA1045, 7CTA0333, 8CTA1044, 8CTA1043, 8CTA0995, 8CTA1024, 8CTA1025, 8CTA1026, 8CTA1027, 8CTA1028, 8CTA1029, 8CTA1030, 8CTA1031, 8CTA1034, 8CTA1035, 8CTA1036, 8CTA1038

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1209-2019
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Fenwal Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1209-2019. Report a data issue: hello@regdatalab.com.