Z-1344-2021 Class II Terminated

Recalled by Fenwal Inc — Lake Zurich, IL

FDA device recall Z-1344-2021 was initiated by Fenwal Inc on Feb 25, 2021 and is designated Class II. Reason for recall: Some lots of Amicus MNC Apheresis and Amicus Exchange kits have leaking centrifuge packs during a procedure on the Amicus Separator using therapeutic protocols only. The defect is a blood leak at… The recall status is terminated (terminated Apr 18, 2023). Affected quantity: 2466 units.

Recall details

Product Type
Devices
Report Date
Apr 14, 2021
Initiation Date
Feb 25, 2021
Termination Date
Apr 18, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2466 units

Product description

Amicus Exchange Kit - Product Usage: intended for use in the collection of blood components and mononuclear cells.

Reason for recall

Some lots of Amicus MNC Apheresis and Amicus Exchange kits have leaking centrifuge packs during a procedure on the Amicus Separator using therapeutic protocols only. The defect is a blood leak at the boot, elbow, or in the channel on the separation chamber of the centrifuge pack.

Distribution pattern

US Nationwide distribution in the states of GA, TX, TN, MD, CA, KS, MN, MA, NM, NC, LA, CO, KY, NV, PA, UT, MI, NJ, IL, MD, OH, NY, FL.

Code information

Product Code X6R2339; UDI: 04086000100151; Batch Numbers FA20F08147 (Exp. 06/07/2022), FA20H12145 (Exp. 08/11/2022), FA20J27157 (Exp. 10/26/2022)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1344-2021
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Fenwal Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1344-2021. Report a data issue: hello@regdatalab.com.